Chapter 6. Calibrator Tasks
LOW
value - is the first current source value to be set using this option.
HIGH
value - is the maximum current source value set-point using this option.
DWELL
- the time interval which can be set to wait at the low value before changing to the high
value and vice versa.
SCALING
- refer to Section 6.3.6 on page 122 for a description of the use of Scaling.
When
Span Check
automation is selected, the
Span Check
icon is displayed on the top right
of the
CH2
window on the calibrator screen.
The span check can be run manually by using the
UP
and
DOWN
nudge buttons to switch
between the
LOW
and
HIGH
source values. It can also be automated by pressing the automation
Play
softkey.
The
Auto-repeat
icon is also displayed next to the
Span Check
automation
icon, as the
Span Check
cycle is automatically repeated when using the automation
Play
softkey, until
manually stopped. To stop the automation at any time, press the
Cancel
softkey.
Step 3
The
Calibrator
screen will have two
additional icons indicating that the
Span Check
mode is active and the
Auto-repeat
operation is available.