Chapter 11. Documenting
Step 4
Cont’d
OUTPUT
: The Output refers to the output
signal from the Device Under Test (DUT).
By default, the
FUNCTION
(and hence
SENSOR
type) set up in
CH2
is configured
as the output with the selected
UNITS
.
These are read-only and not editable in the
procedure creation wizard. If any changes
are required, these must be done in
Calibrator prior to using the Documenting
application.
The output range is auto populated by
default with the full sensor range related to
the Function. Only the range can be edited
and must be within the full sensor range.
Set the
START
and
END
values of the input
range if different to the values dis-played.
Note
: It is possible using the Toggle softkey
to swap the Input and Output functions i.e.,
use the original Input function as the Output
function and vice versa.
TEST POINTS
: This refers to input points at
which the device under test (DUT) will be
tested and its calibration data recorded.
Points Up & Points Down
The number of calibration points must be
specified within the specified Input range.
This can be specified as
POINTS UP
i.e.
direction from
START
range value to
END
range value or/and
POINTS DOWN
i.e.
direction from
END
range value to
START
range value.
The default setting is 2 Points Up and 0
Points Down. This means there will be two
test points - the first test point will be the
Input Start value and the second will be the
Input End value.