English–DPI610E Instruction Manual | 139
TX (Transmitter Simulation) Simulator
Step 3
The
Calibrator
screen will be
configured with the transmitter
simulation details.
CH1
will be
configured with the internal pressure
function while
CH2
will be configured
with
Current
source in simulation
mode. The
TX Sim
icon will be
displayed within the
CH2
window under
the
Function
name field.
Note
: In
TX Sim
mode, the current
output is automatically determined, and
source based on the transfer function
characteristic configured.
Step 4
To configure the simulated transmitter:
Tap in the
CH2
area to maximize the
CH2
window and select the
Setup
softkey.
Select and edit the
START
and
END
values of the Input channel i.e. the
Pressure
channel. Default values are
the zero (or Negative full-scale) and the
Positive full-scale of the internal
pressure sensor respectively.
Select and edit the
Start
and
End
values of the
OUTPUT
channel i.e. the
current source channel. Default values
are 0 and 24 mA.
There is also the option for the
DPI 610E to provide power i.e. 10V or
24V. The default setting is
Off
(no
power provided by the DPI 610E).
Press the
Back
softkey.