34
Parameter Setup Mode
The 2200 series provides many parameters that allow you to customize the operation of
your 2200 indicator to meet your application’s needs. To access these parameters the
setup mode must be accessed, which can be entered using any of the three methods
below.
Power–up Front Panel Setup Mode Access
To enter the calibration mode, power up the indicator while pressing and holding the
ZERO and the UNITS buttons. When
rEL Pb
is displayed, release both buttons. The
display will momentarily read
Ent Cd
, and then go blank. Press the ZERO button five
times. The display will indicate the number of times the ZERO button has been pressed.
When
5
is displayed, press the UNITS button and wait a few seconds.
Note:
If the code is not entered before the timer is finished, the scale will bypass code
entry and enter the normal run mode. The front panel access feature during power-up is
not available when the Operating Mode (
oP
) parameter is set to
44S
.
Front Panel Setup Mode Access
To enter the calibration mode, press and hold the UNITS and ZERO buttons until the
parameter review starts (
C
and
P
are displayed). Press the HIDDEN (Capacity Label)
button after
Cap aj
and the capacity is displayed. The display will momentarily read
Ent Cd
, and then go blank. Press the ZERO button five times. The display will indicate
the number of times the ZERO button has been pressed. When
5
is displayed, press
the UNITS button and wait a few seconds.
Note:
If the code is not entered before the timer is finished, the scale will bypass code
entry and enter the normal run mode. The front panel access feature during power-up is
not available when the Operating Mode (
oP
) parameter is set to
44S
.
Switch Setup Mode Access
The calibration switch can be accessed by removing the meter’s back cover. With the
indicator powered on, press the CAL switch (S1), located in the lower left corner of the
main board (see Fig. 6). Pressing the CAL switch also exits the setup mode and saves
any changes.
Warning:
do not press the CAL switch while powering up the scale, this will cause the
scale to reset all parameter settings.
Exiting the Setup Mode
To exit the Calibration and Parameter Setup Menu, momentarily press the CAL switch or
scroll through the menu options, by pressing the UNITS button, until
donE n
appears.
Press the ZERO button until
donE y
appears and then press the UNITS button. The
indicator will return to the normal weighing mode. If any menu selections were changed,
the new values will be saved.
Note:
No new setup information is saved until the scale displays
SAVEd
and returns to
the weigh mode. In the event of a power failure while in the setup mode, any changes
that have been made will be lost.