52
Test Mode Menu
XXXXXX
Displays the raw counts from Analog to Digital converter. Press the
Zero button to enter Test mode menu
4nnA
Set 4-20mA output to 4mA. Press ZERO to change the output level.
4A
XX
Press the ZERO and PRINT pushbuttons to adjust 4mA
level and UNITS pushbutton to exit. (default value = 97)
20nnA
Set 4-20mA output to 20mA. Press ZERO to change the output level.
20A
XX
Press ZERO and PRINT pushbutton to adjust 20mA level
and UNITS pushbutton to exit. (default value = 3C)
out1
Activate Output 1 at TB4 terminal. Press ZERO to select the type of logic
options for Output 1. Use UNITS to scroll to the next Output parameter.
o1 oFf
No output logic, Output 1 is turned off.
o1 SP1
Setpoint 1 used for output logic. (default)
o1 SP2
Setpoint 2 used for output logic.
o1 SP3
Setpoint 3 used for output logic.
o1 SP4
Setpoint 4 used for output logic.
o1 SP5
Setpoint 5 used for output logic.
o1 SP6
Setpoint 6 used for output logic.
o1 SP7
Setpoint 7 used for output logic.
o1 SP8
Setpoint 8 used for output logic.
o1 Er
Scale Error Message used for output logic.
o1 THr
Weight below threshold level used for output logic.
o1 in1
Remote Switch Input Logic 1 used for output logic.
o1 in2
Remote Switch Input Logic 2 used for output logic.
o1 bAt
Output controlled by Batching operation. (Model 2200B)
o1 Lo
Low used for output logic. (Model 2200CW)
o1 udr
Under used for output logic. (Model 2200CW)
o1 Acc
Accept used for output logic. (Model 2200CW)
o1 our
Over used for output logic. (Model 2200CW)
o1 Hi
High used for output logic. (Model 2200CW)
o1 SEr
Serial port used for output logic.
o1 Alt
Alarm countdown timer used for output logic. (
oP
=
ALt
)
out 2
Activate Output 2 at TB4. Press Zero pb to select output logic.
o1 SP2
Setpoint 2 used for output logic. (default)
out 3
Activate Output 3 at TB4. Press Zero pb to select output logic.
o1 SP3
Setpoint 3 used for output logic. (default)
out 4
Activate Output 4 at TB4. Press Zero pb to select output logic.
o1 SP4
Setpoint 4 used for output logic. (2200, 2200B default)
o1 Lo
Low used for output logic. (2200CW default)
out 5
Activate Output 5 at TB4. Press Zero pb to select output logic.
o1 SP5
Setpoint 5 used for output logic. (2200, 2200B default)
o1 udr
Under used for output logic. (2200CW default)
out 6
Activate Output 6 at TB4. Press Zero pb to select output logic.
o1 SP6
Setpoint 6 used for output logic. (2200, 2200B default)
o1 Acc
Accept used for output logic. (2200CW default)
out 7
Activate Output 7 at TB4. Press Zero pb to select output logic.
o1 SP7
Setpoint 7 used for output logic. (2200, 2200B default)