Defaulting Wired and Wireless Ethernet Modules
The wireless and wired Ethernet modules can be reset to factory defaults by shorting the
two-pin header JU42 with the supplied jumper. This should be done with the unit
powered
off.
With the jumper in place, power the unit on. Once the amber (wireless) or
green (wired) LED begins blinking a 1-5-1 pattern, remove the jumper. The module is
now successfully defaulted. After about one minute the module will attempt to make a
connection to a WAP that is broadcasting its SSID and has security disabled. Once
defaulted, the module needs to be configured for proper network communication.
Run the
dgdiscvr.exe
software provided with the Doran Scale on a PC connected to the
network. When this program is executed, all Doran Wired and Wireless Ethernet
devices connected to your network will be listed.
Highlight the device to be modified and click on “Open web interface”. If you have
multiple devices that are new to the network that need to be configured, the scale can be
identified by the MAC address. The MAC address can be found just above the antenna
on the rear of the scale.
A web browser will be launched that will allow you to reconfigure the selected device as
seen below.