76
[INUA] Request of a numeric value entry on the 3590E LCD display / CPWE-CPWET data section:
Syntax
Format
<CC>INUA<PP><LL><I><CR LF>
Parameters
PP = decimal position from which the entry of the numeric data starts,
from left to right.
LL = number of characters in decimals which make up the numeric data
to be entered
I = selection/unselection of the display of an initial value, previously
entered in the user buffer (using the WUBU command described
later on):
- I = 0 not displayed
- I = 1 displayed
Answer
<CC>OK<CR LF>
Example
01INUA08071<CR LF>
Result
If preceded by the 01WUBU1234567<CR LF> command, it causes the
1234567 message to be shown on the 3590E LCD display / CPWE-CPWET
data section, with the character 1 blinking, starting from the position 08
of the 3590E LCD display / CPWE-CPWET data section (in other words
the digit 8). At this point it is possible to set 8 numerical digits which will
substitute the displayed ones.
NOTE
: If an alphanumeric string is in the buffer, a number of blank spaces corresponding to the number of digits set in LL
will be shown on the display, starting from the set cursor position.
[IALA] Request of alphanumeric text entry on the 3590E LCD display / CPWE-CPWET data section:
Syntax
Format
<CC>IALA<PP><LL><I><CR LF>
Parameters
PP = decimal position from which the entry of the numeric data starts,
from left to right.
LL = number of characters in decimals which make up the numeric data
to be entered
I = selection/unselection of the display of an initial value, previously
entered in the user buffer (using the WUBU command described
later on):
- I = 0 not displayed
- I = 1 displayed
Answer
<CC>OK<CR LF>
Example
01IALA08071<CR LF>
Result
If preceded by the 01WUBUABCDEFG<CR LF> command, it causes the
ABCDEFG message to be shown on the 3590E LCD display / CPWE-
CPWET data section, with the character 1 blinking, starting from the
position 08 of the 3590E LCD display / CPWE-CPWET data section (in
other words the digit 8). At this point it is possible to set 8 numerical
digits which will substitute the displayed ones.
When exiting the entry phase, the displayed data is frozen on the 3590E LCD display / CPWE-CPWET data section and it
remains until a new serial command does not switch the visualisations (for example a DISP command or a new
visualisation entered with the INUA or IALA commands, previously described).
The turning off causes the restoration of the standard weight data.