3590EKR, 3590EXP, 3590EXT, 3590 EBOX, CPWE, CPWET series indicator E-AF03_05.01_14.07_EN_T.doc
58
3.3.8 THEORETICAL CALIBRATION
Premise:
it’s possible to use this procedure if one does not have a sample weight available for carrying out a real
calibration.
1)
Follow the steps described in section 3.3.1
up to point 15;
2)
Select the
SEtuP >> ConFiG >> thEo.CA
step and press
ENTER
;
3)
It will be requested if proceed with the new calibration ("SurE?"). Press
ENTER
to confirm, or
C
to cancel.
4)
Set the cell capacity
Enter the cell capacity (the unit of measure is the one configured for the scale: minimum value scale capacity,
maximum value 999999) and confirm with
ENTER
.
5)
In case of analogue cells:
Set the cell sensivity in mV/V :
Enter the cell sensivity in mV/V (max 99.99999 mV/V) and confirm with
ENTER
.
In case of various connected cells through the junction box, enter the average of the value; in case of
dependent channels, enter the sum of the value.
In case of digital cells:
Set the cell points at the full range capacity :
Enter the full range capacity digital value (see the load cell datasheet: minimum value is double the scale
capacity, maximum value 9999999) and confirm with
ENTER
.
At the end of the calibration, the display shows the message "CALIBRATION EXECUTED" and automatically exits the step,
applying the new calibration.