the non significant digits (for example, if the transmitted string is
spp CR + LF
, in which
s
is the weight
sign,
ppppppp
is the weight value,
u
is the unit of measure, one should set the value 08).
(!) 01
<< W.tYPE>>
WEIGHT TYPE
In this step it’s possible to choose whether the previously configured weight value is a gross weight or a net weight:
GroSS
Gross weight
nEt
Net weight
(!) GroSS
NOTE:
The following tare settings are not necessary if the remote scale transmits a string containing both the gross
and the tare weights.
<< tAr.PoS >>
TARE WEIGHT POSITION
In this step one sets the position of the first character of the tare value in the string transmitted from the remote
scale, taking into account that the first character on the left of the string has position 00. It’s possible to set up to 3
characters (from 000 to 100); by setting 255, the reading of value parameter is disabled.
(!) 255
<< tAr.LEn >>
LENGTH OF THE TARE WEIGHT STRING
In this step one enters the number of digits (from 1 to 99) which make up the tare value, including the non
significant digits and the decimal point.
(!) 01
<< tAr.tYP >>
TARE TYPE POSITION
In this step one sets the position of the first character of the tare type indication (2 characters), in the string
transmitted from the remote scale, knowing that the first character on the left of the string has the position 00; it’s
possible to set up to 3 characters (from 000 to 100); by setting 255, the reading of the preset tare indication is
disabled.
If the value in the indication corresponds to “PT” the previously configured tare is considered as preset; otherwise it
is considered as a semiautomatic tare.
(!) 255
EXAMPLE:
if the transmitted string is
spppppppuu, ttttttttuu kk + CR + LF
, in which
tttttttt
is the tare value,
uu
is the unit of
measure and
kk
is the type of tare:
-
tAr.PoS:
12
-
tAr.LEn:
08
-
tAr.tYP:
23
<< Str.LEn >>
LENGTH OF THE COMPLETE STRING
In this step one enters the number of digits (from 0 to 99) which make up the entire string transmitted from the scale
serial line minus the terminator character (for example, if the transmitted string is
spp CR + LF
, in which
s
is the weight sign,
ppppppp
is the weight value,
u
is the unit of measure, one should set the value 10).
Configure 0 in case of variable weight data string length; the position of the weight characters in the data string can
not change.
(!) 00
<< CAPAC. >>
CAPACITY OF THE REMOTE SCALE
Enter the maximum capacity (up to 6 characters, from 1 to 999999) of the remote scale, taking into account the
number of decimals of the scale.
(!) 000001
<< dIV. >>
REMOTE SCALE’S DIVISION
Enter the scale’s division (up to 3 characters, from 1 to 200). If it functions in dual or triple range, enter the division
of the lower range.