3590EKR, 3590EXP, 3590EXT, 3590 EBOX, CPWE, CPWET series indicator E-AF03_05.01_14.07_EN_T.doc
71
Reading of converter points relative to the weight:
[CC]RAZF<CR LF>
Instrument’s answer:
[CC]SS,RZ,RRRRRRRRRR,vv<CR LF>
In which:
SS
UL
Underload
OL
Overload
ST
Stability of the weight
US
Instability of the weight
RRRRRRRRRR Value of converter points (10 characters)
Reading of microvolts relative to the weight: [CC]MVOL<CR LF>
Instrument’s answer:
[CC]SS,VL,MMMMMMMMMM,uv<CR LF>
In which:
SS
UL
Underload
OL
Overload
ST
Stability of the weight
US
Instability of the weight
MMMMMMMMMM Value of microvolts (10 characters)
Reading of net weight with sensitivity times 10: [CC]GR10<CR LF>
Instrument’s answer:
[CC]SS,GX,VVVVVVVVVV,UM<CR LF>
In which:
SS
UL
Underload
OL
Overload
ST
Stability of the weight
US
Instability of the weight
GX
Weighing times 10 status
VVVVVVVVVV Value of net weight times 10 (10 characters)
UM
Unit of measure (Kg, g, t, lb)
NOTE
: The instrument does not transmit the OK answer to the short commands (R, T, Z, P….).
Modification of GR10 command response: [CC]GR10X<CR LF>
in which:
GR10= command
X
E
enabled
X
D
disabled
Example: GR10E<CR LF>
If enabled, it modifies the format of the GR10 command response string: it responds to the GR10 command with the
number of the active scale in the place of the “GX – weighing status x 10” ).
The setting is valid when the indicator is turned off. To save it permanently in the instrument one should transmit the
command in the set-up status.
Setpoint command:
[CC]STPTnFxxxxxxOyyyyyy<CR LF>
in which:
n, expressed in hexadecimals, indicates the number of SETPOINTS (1, 2, 3, 8, 9,
A, B, C, D, E, F).
F
: the following weight value indicates the DISABLING of the relays (OFF).
O
: the following weight value indicates the ENABLING of the relays (ON).
xxxxxx e yyyyyy take on the enabling or disabling setpoint value: the digits must
be entered WITHOUT decimal point, omitting the NON significant zeros.
Example in case of instrument with 10,000 kg capacity and 1 g division:
Command :
STPT1F5000O6500
(Disabling of first relay at 5 kg and enabling at 6,5 kg)
Answer:
OK
NOTE
: Negative answer in the following conditions:
-
one of the two entered values surpasses the capacity.
-
one of the two entered values has a minimum division which is incongruent with the one set in the instrument.
-
the disabling value surpasses the enabling one.