8 Electromagnetic Conformity
8.1 Guidance and Manufacturer’s Declaration
The essential performance of the RMU-1000 ACC and its accessories is to provide accurate chest
compression depths and rates.
The RMU-1000 ACC is intended for use within the electromagnetic environment specified below.
The customer or the user of the RMU-1000 ACC should assure that it is used within the stated
environmental specifications.
Electromagnetic Emissions
Emissions test
Compliance
Electromagnetic environment – guidance
RF emissions
The RMU-1000 uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic
equipment.
CISPR 11
Group 1
Class B
Harmonic emissions IEC
61000-3-2
Class A
The RMU-1000 is suitable for use in all buildings including
domestic homes and places directly connected to the
public low-voltage Power Supplies Network that supplies
buildings used for domestic purposes.
Voltage fluctuations/flicker
emissions IEC 61000-3-3
Complies
Electromagnetic Immunity
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment – guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±8 kV contact
±15 kV air
There are no special requirements with
respect to electrostatic discharge.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power line
supply lines
±1 kV for input/output
lines
±2 kV for power line
supply lines
±1 kV for input/output
lines
The mains power quality must be that of a
typical commercial or hospital environment.
Surge
IEC 61000-4-5
±2 kV
±2 kV
The mains power quality must be that of a
typical commercial or hospital environment.
Voltage dips,
short interruptions
and voltage
variations on
power supply
input lines IEC
61000-4-11
Ut = 0%, 0.5 cycle
Ut = 0%, 1 cycle
Ut = 70%, 25/30 cycles
Ut = 0%, 250/300 cycles
Ut = 0%, 0.5 cycle
Ut = 0%, 1 cycle
Ut = 70%, 25/30 cycles
Ut = 0%, 250/300 cycles
The mains power quality must be that of a
typical commercial or hospital environment.
If the user of the [Equipment or System]
requires continued operation during
power mains interruptions, Defibtech
recommends that the [Equipment or
System] is energized from a Power Supply
or Battery Pack that cannot be interrupted.
Power frequency
(50/60 Hz)
magnetic field IEC
61000-4-8
30 A/m
30 A/m
Power frequency magnetic fields should
not be greater than levels characteristic of a
typical location in a commercial or hospital
environment.
8. Electr
omagnetic
Conf
ormity
RAC-E1510EN-DN
55
Содержание RMU-1000
Страница 2: ......
Страница 6: ...RAC E1510EN DN 6...
Страница 8: ...1 2 The Defibtech RMU 1000 ACC 7 6 3 4 5 1 2 13 12 14 15 8 9 10 11 RAC E1510EN DN 8...
Страница 12: ...RAC E1510EN DN 12...
Страница 28: ...RAC E1510EN DN 28...
Страница 50: ...RAC E1510EN DN 50...
Страница 54: ...RAC E1510EN DN 54...
Страница 64: ...RAC E1510EN DN 64...