• The ACC can become hot during extended use. The applied part of the piston may rise 5°C
above ambient.
• If the position of the Piston changes as a result of defibrillation or other therapies,
immediately stop compressions and re-adjust the position of the ACC.
• Mechanical chest compressions may cause artifact and interfere with ECG analysis.
Always pause compressions before performing ECG analysis with other equipment.
• When the Battery Pack indicator shows one red segment, replace the Battery Pack as
soon as possible with a sufficiently charged Battery Pack or apply external power.
• To avoid the risk of electric shock, this equipment must only be connected to a supply
mains with protective earth.
• If a spare Battery Pack or external power source are not available and the ACC stops
compressions, remove the unit from the patient and begin manual compressions
immediately.
• If there is a malfunction, the compressions are not sufficient, or something unusual
occurs during operation, then push the ON/OFF Button for one second to stop the ACC
from delivering compressions and remove the unit from the patient. Start manual chest
compressions as soon as possible.
• If there is a malfunction and the unit will not turn OFF, remove Battery Pack to stop
compressions. Remove the unit from the patient. Start manual chest compressions as
soon as possible.
• The Patient Wrist Straps are designed only for use with Defibtech RMU-1000 Automated
Chest Compression Systems. Do not apply to other items such as a stretcher or bedside
rails.
• Do not use the Patient Wrist Straps if damaged, soiled, or if Velcro does not bond
sufficiently to securely fasten the patient’s wrists to the RMU-1000 ACC.
• Do not over-tighten the Patient Wrist Straps to prevent potential injury. If the Patient
Wrist Straps are too loose, they may not securely fasten the patient’s arms to the ACC
appropriately.
• Do not use the Patient Wrist Straps to lift the patient. The straps are only intended to
attach the patient’s arms to the RMU-1000 ACC during transport.
• Observe the patient’s arms during use of the Patient Wrist Straps. Do not leave a patient
unattended when using the ACC or its accessories (e.g. Patient Wrist Straps).
• Do not obstruct intravenous (IV) access when deploying the RMU-1000 or when attaching
the Patient Wrist Straps.
• Use of damaged equipment or accessories may cause the device to perform improperly
and/or result in injury to the patient or operator.
• The ACC system components must be cleaned and the Patient Interface Pad replaced
between patients to avoid cross contamination.
• The use of accessories other than those specified may result in increased emissions or
decreased immunity of the RMU-1000 ACC.
• Portable and mobile RF communications equipment should be used no closer to any part
of the RMU-1000 ACC, including cables, than necessary.
WARNINGS (continued)
RAC-E1510EN-DN
14
Содержание RMU-1000
Страница 2: ......
Страница 6: ...RAC E1510EN DN 6...
Страница 8: ...1 2 The Defibtech RMU 1000 ACC 7 6 3 4 5 1 2 13 12 14 15 8 9 10 11 RAC E1510EN DN 8...
Страница 12: ...RAC E1510EN DN 12...
Страница 28: ...RAC E1510EN DN 28...
Страница 50: ...RAC E1510EN DN 50...
Страница 54: ...RAC E1510EN DN 54...
Страница 64: ...RAC E1510EN DN 64...