4.7 Power
If the Battery Pack charge becomes low during use, the warning and Battery Pack indicators will
flash. The Battery Pack status indicator will show only one red indicator bar.
OPTION 1)
If a charged spare Battery Pack is available:
• Obtain the spare Battery Pack and have it ready for installation.
• Push Pause to temporarily stop compressions.
• Quickly eject the Battery Pack from the ACC by pressing the Battery Pack Release and
remove the depleted Battery Pack.
• With minimal interruption, install the spare Battery Pack.
• Wait for the Pause LED indicator to illuminate.
• Re-start compressions by pushing the Pause button again or one of the Run buttons.
Note:
If the Battery Pack change takes more than approximately 15 seconds, the unit will power
off with the piston in place. Upon spare Battery Pack insertion, the Compression Module must
be powered on and the piston will automatically retract to set its start position. The piston should
then be re-adjusted to patient’s chest.
OPTION 2)
At any time, the ACC can be connected to an external power source by connecting
the AC adapter to the external input jack of the Compression Module. See Section 3.9 for details.
RAC-E1510EN-DN
40