7.2 Battery Pack
Use only Defibtech Battery Packs in the RMU-1000 ACC.
RBP-1000 Battery Pack
Category
Specification
Model number
RBP-1000
Battery type
18.5V, 5300 mAh, Lithium-ion.
Rechargeable, recyclable.
Operation time
1 hour (nominal patient)*
Battery Pack charge time
Less than 3 hours in ACC*
Less than 2 hours if charging one Battery Pack in optional external battery
pack charging station (less than 3 hours if charging two Battery Packs;
see Section 6.7 for details)*
Battery Pack useful life
Recommended to replace Battery Pack every 3 years or if Battery
Pack indicator displays a replace Battery Pack condition (~300 charge/
discharge cycles**).
Battery Pack operating temperature
0 – 40°C (32 – 104°F) ambient
Charging temperature
0 – 40°C (32 – 104°F) ambient
Storage temperature
0 – 40°C (32 – 104°F);
-20 – 60°C (-4 – 140°F) short-term <1 month.
Sealing / water resistance
IEC 60529 class IP44
*typical, new battery, at 25°C
**one charge/discharge cycle is defined as charging
and discharging the full capacity of the Battery Pack
7.3 AC Power Adapter
RPM-1000 External AC Power Adapter
Category
Specification
Model number
RPM-1000
Input voltage
100 – 240VAC, 50/60Hz nominal
Input current
≤2.3A
Rated output
24.0VDC (±5%)
Operating / storage temperature
See
“Environmental”
specifications chart in Section 7.1
Emissions and immunity
Refer to Chapter 8 for details
RAC-E1510EN-DN
52
Содержание RMU-1000
Страница 2: ......
Страница 6: ...RAC E1510EN DN 6...
Страница 8: ...1 2 The Defibtech RMU 1000 ACC 7 6 3 4 5 1 2 13 12 14 15 8 9 10 11 RAC E1510EN DN 8...
Страница 12: ...RAC E1510EN DN 12...
Страница 28: ...RAC E1510EN DN 28...
Страница 50: ...RAC E1510EN DN 50...
Страница 54: ...RAC E1510EN DN 54...
Страница 64: ...RAC E1510EN DN 64...