4.9 Removal from Patient
To remove the ACC from the patient:
• Turn the ACC off by pressing and holding the ON/OFF button for at least one second.
• To release a patient’s arm from the Patient Wrist Strap, pull up on the fabric tab on the end
of the Forearm Strap until the Velcro is no longer holding the patient’s arm in place. Repeat
for the patient’s other arm.
• Remove the Stabilization Strap (see Section 4.4).
• Press down and rotate the Compression Module approximately 90 degrees in either
direction. Lift Compression Module from Frame. (
Note:
This step may be performed after
Frame is removed from patient.)
• Press the latch release levers to disconnect the Frame from the Backboard.
• Remove the Backboard from underneath the patient.
4.10 Post-Use Procedures
After the RMU-1000 ACC has been used on a patient, the unit should be cleaned following
procedures in the “Cleaning” section in Chapter 5 of this manual and prepared for the next use.
The following steps should be performed:
• Remove and dispose of the used Patient Interface Pad.
• Remove and clean the Stabilization Strap.
• The Patient Wrist Straps may be removed for cleaning purposes or if they need to be
replaced (see Section 5.2). To remove the Patient Wrist Straps from the ACC’s Frame, pull
up on the rounded end of the Frame Strap until it can be slid through the Frame Strap Loop.
Note:
To maximize the available time to perform a rescue, Defibtech recommends that the
Patient Wrist Straps remain affixed to the ACC’s Frame and that the Frame be stored in the
ACC’s Carrying Case with the Patient Wrist Straps already attached to the Frame.
• Clean all the components that have been in contact with the patient and let them dry (see
Section 5.2).
• Replace the Battery Pack with a fully charged Battery Pack or fully charge the Battery Pack
in the ACC unit.
• Install a new Patient Interface Pad.
• Repack the ACC components and spares in the Carrying Case.
4.11 Operational Environment
The Defibtech ACC is designed to operate in a wide range of environmental conditions. To ensure
the reliability and safety of the ACC in a given environment, refer to the “Environmental” section
in Chapter 7 of this manual for a detailed list of specified environmental conditions.
RAC-E1510EN-DN
42
Содержание RMU-1000
Страница 2: ......
Страница 6: ...RAC E1510EN DN 6...
Страница 8: ...1 2 The Defibtech RMU 1000 ACC 7 6 3 4 5 1 2 13 12 14 15 8 9 10 11 RAC E1510EN DN 8...
Страница 12: ...RAC E1510EN DN 12...
Страница 28: ...RAC E1510EN DN 28...
Страница 50: ...RAC E1510EN DN 50...
Страница 54: ...RAC E1510EN DN 54...
Страница 64: ...RAC E1510EN DN 64...