3.7 Installing and Removing the Patient Interface Pad
Attach a Patient Interface Pad to the distal end of the Piston by pressing the pad onto the piston
until it snaps into place, rotating pad if necessary. To remove the Patient Interface Pad, grasp the
pad and pull, as shown below.
Note:
Prior to installation, check to make sure that the Patient Interface Pad is clean and is
undamaged. If the Patient Interface Pad is damaged (e.g. one or more of the pins on underside of
the pad is crushed and/or missing), discard it and install an undamaged pad.
Installing
Removing
3.8 Installing and Removing the Battery Pack
The Battery Pack provides power to the RMU-1000 ACC. Do not install the Battery Pack after the
expiration date printed on the label.
Before inserting the Battery Pack into the RMU-1000 ACC, ensure
that the Battery Pack opening in the side of the Compression Module
is clean and clear of any foreign objects. Insert the Battery Pack into
the opening on the side of the ACC Compression Module. Push the
Battery Pack all the way in until the latch clicks. The Battery Pack will
operate in either orientation (with the contacts toward the unit).
When the Battery Pack is inserted, the RMU-1000 ACC will display the status of the Battery Pack
on the Compression Module's Battery Pack indicator for approximately three seconds.
Ship Mode:
Battery Packs are shipped from the factory at less than 30% charge capacity
in a low-power state known as ship mode to minimize energy loss during shipment. Any
Battery Pack in ship mode cannot power the ACC until it is taken out of ship mode and
should be transferred from ship mode to active mode as soon as possible upon receipt.
To transfer a Battery Pack from ship mode to active mode, insert the Battery Pack into the
Compression Module as described earlier in this section. Then connect the AC Adapter to
the Compression Module, as described in Section 3.9,
“Charging the Battery Pack.”
Upon
detecting external power (allow at least 10 seconds for this to occur), the Battery Pack will
automatically switch from ship mode to active mode. The optional Battery Pack Charging
Station (see Section 6.7) can also be used to switch Battery Packs from ship mode to
active mode. Defibtech recommends charging any Battery Pack that has been brought out
of ship mode to full capacity (see Section 3.9 for details). In addition, a Battery Pack should
not be stored in ship mode for any longer than 12 months before charging it to full capacity.
CAUTION
A new battery is set to ship mode. Must be activated before use!
Activate battery by charging in unit or charger.
3.
S
et
ting Up the
RMU-1
0
0
0 A
CC
RAC-E1510EN-DN
23
Содержание RMU-1000
Страница 2: ......
Страница 6: ...RAC E1510EN DN 6...
Страница 8: ...1 2 The Defibtech RMU 1000 ACC 7 6 3 4 5 1 2 13 12 14 15 8 9 10 11 RAC E1510EN DN 8...
Страница 12: ...RAC E1510EN DN 12...
Страница 28: ...RAC E1510EN DN 28...
Страница 50: ...RAC E1510EN DN 50...
Страница 54: ...RAC E1510EN DN 54...
Страница 64: ...RAC E1510EN DN 64...