5 Maintenance and Troubleshooting
This chapter describes the maintenance and troubleshooting procedures for the RMU-1000 ACC.
The unit warnings and alerts are described along with recommended routine maintenance. A
troubleshooting guide is provided to help diagnose user serviceable problems.
The RMU-1000 ACC contains no user serviceable parts.
5.1 Routine Unit Maintenance
The RMU-1000 ACC is designed to be very low maintenance. Simple maintenance tasks are to
be performed regularly to ensure its readiness (see sample maintenance table below). Different
maintenance intervals may be appropriate depending on the environment where the RMU-1000
ACC is deployed, and ultimately the maintenance program is at the discretion of the emergency
response program's medical director. As with all rechargeable batteries, the rechargeable Battery
Pack self-discharges over time, and therefore, must be maintained in accordance with the User
Manual in order to be ready for use.
Weekly
After Each Use
Action
•
•
Make sure the Battery Pack is fully charged (see Section 3.9, “Charging
the Battery Pack”).
•
•
Check the condition of the system. Make sure the Carrying Case contains
accessories including Patient Interface Pad(s), AC Adapter, Stabilization
Strap, and Patient Wrist Straps.
•
•
Make sure a Patient Interface Pad is installed on the Compression
Module.
•
•
Make sure at least one unused Patient Interface Pad is stored in the
Carrying Case.
•
Check the Battery Pack expiration date.
•
•
Turn the ACC on to perform a self-test. Make sure the piston is retracted
and the PAUSE indicator comes on with no warning indicators.
Note:
If the unit has been dropped, mishandled, or abused, a thorough evaluation of operation
should be performed.
5.2 Cleaning
After each use, clean any dirt or contaminants from the RMU-1000 ACC’s Frame, Backboard,
and Compression Module. The following are important guidelines that must be adhered to when
cleaning the device:
• To clean the RMU-1000 ACC’s Frame, Backboard, or Compression Module, use a soft
cloth dampened with one of the following recommended cleaning agents:
– Soapy water
– Ammonia-based cleaners
– Hydrogen peroxide
– Isopropyl alcohol (70 percent solution)
– Chlorine bleach (30 ml/liter water)
• Do not immerse the RMU-1000 ACC’s Frame, Backboard, or Compression Module,
components in fluids or allow fluids to enter the unit.
• Do not spray cleaning solutions directly on the unit or its connectors.
• Do not use abrasive materials or strong solvents such as acetone or acetone-based
cleaning agents.
5.
Maintenance and
Tr
oubleshooting
RAC-E1510EN-DN
43
Содержание RMU-1000
Страница 2: ......
Страница 6: ...RAC E1510EN DN 6...
Страница 8: ...1 2 The Defibtech RMU 1000 ACC 7 6 3 4 5 1 2 13 12 14 15 8 9 10 11 RAC E1510EN DN 8...
Страница 12: ...RAC E1510EN DN 12...
Страница 28: ...RAC E1510EN DN 28...
Страница 50: ...RAC E1510EN DN 50...
Страница 54: ...RAC E1510EN DN 54...
Страница 64: ...RAC E1510EN DN 64...