2 Warnings and Cautions
This chapter includes a list of warnings and cautions that relate to the RMU-1000 ACC and its
accessories. Many of these messages are repeated elsewhere in this User Manual and on the
RMU-1000 ACC or accessories.
2.2
WARNINGS:
Immediate hazards that will result in serious personal injury or death.
None known.
Conditions, hazards, or unsafe practices that may result in serious personal injury or death.
• Improper use can cause injury to patient. Use the RMU-1000 ACC only as instructed in the
User Manual.
• Improper use can cause injury to operator or bystander. Keep fingers and hands away from
Piston during operation.
• Improper maintenance can cause the RMU-1000 ACC not to function. Maintain the RMU-
1000 ACC and rechargeable Battery Pack only as described in the User Manual. Failure to
maintain the Battery Pack per the instructions outlined in this User Manual will result in the
RMU-1000 ACC becoming inoperable.
• No modification of this equipment is allowed. The RMU-1000 ACC contains no user-
serviceable parts. Do not disassemble, repair or alter the RMU-1000 ACC or any of its
components.
• Do not immerse Compression Module or Battery Pack in water or other liquids. Immersion
in fluids may result in fire or explosion.
• Do not sterilize the RMU-1000 ACC or its accessories.
• Do not let fluids get into the RMU-1000 Compression Module. Avoid spilling fluids on the
ACC or its accessories. Spilling fluids into the RMU-1000 ACC may damage it or cause a
fire or shock hazard.
• If the patient is too large for the Frame, remove Frame and continue manual CPR
compressions. Do not use the ACC if the Frame cannot be latched to the backboard.
• If the Piston cannot be adjusted to reach the patient's chest, the patient is too small.
Remove Frame and continue with manual CPR compressions.
• Do not initiate ACC compressions if the Piston is not in the proper position. An incorrect
start position may compromise the patient’s blood circulation.
• Do not use the ACC if the Frame cannot be latched in place (for any reason).
• Incorrect position over chest can result in injury or lack of effectiveness.
• Incorrect Piston start height can result in injury or lack of effectiveness.
• Carefully monitor the position of the Piston on the patient's chest to ensure that it has not
moved from the appropriate target area.
• Changed position over the chest during operation can result in injury or lack of effectiveness.
• Do not leave the ACC running while unattended. Patient injury may result if the unit is left
unattended.
2.
W
arnings and Cautions
RAC-E1510EN-DN
13
Содержание RMU-1000
Страница 2: ......
Страница 6: ...RAC E1510EN DN 6...
Страница 8: ...1 2 The Defibtech RMU 1000 ACC 7 6 3 4 5 1 2 13 12 14 15 8 9 10 11 RAC E1510EN DN 8...
Страница 12: ...RAC E1510EN DN 12...
Страница 28: ...RAC E1510EN DN 28...
Страница 50: ...RAC E1510EN DN 50...
Страница 54: ...RAC E1510EN DN 54...
Страница 64: ...RAC E1510EN DN 64...