WARNING
Incorrect position of the Piston over chest can result in injury or lack of
effectiveness.
WARNING
Incorrect Piston start height can result in injury or lack of effectiveness.
WARNING
Carefully monitor the position of the Piston on the patient’s chest to ensure
that it has not moved from the appropriate target area.
WARNING
Changed position over the chest during operation can result in injury or lack
of effectiveness.
WARNING
ACC compressions may interfere with ECG analysis. Pause compressions
during ECG analysis.
WARNING
When Battery Pack indicator shows one red segment, replace Battery Pack
as soon as possible or apply external power.
WARNING
Do not leave the ACC running while unattended. Patient injury may result if
the unit is left unattended.
WARNING
If there are malfunctions, interruptions, the compressions are not sufficient,
or something unusual occurs during operation, then push the ON/OFF Button
for one second to stop the ACC from delivering compressions and remove the
unit from the patient. Start manual chest compressions as soon as possible.
CAUTION
Improper operation may injure operator or bystander. Keep fingers and hands
away from Piston during operation.
Operation and Adjustment (continued)
4.
Using the
RMU-1
0
0
0 A
CC
RAC-E1510EN-DN
35
Содержание RMU-1000
Страница 2: ......
Страница 6: ...RAC E1510EN DN 6...
Страница 8: ...1 2 The Defibtech RMU 1000 ACC 7 6 3 4 5 1 2 13 12 14 15 8 9 10 11 RAC E1510EN DN 8...
Страница 12: ...RAC E1510EN DN 12...
Страница 28: ...RAC E1510EN DN 28...
Страница 50: ...RAC E1510EN DN 50...
Страница 54: ...RAC E1510EN DN 54...
Страница 64: ...RAC E1510EN DN 64...