• After cleaning, allow the unit to completely dry. Before returning it to service, always turn
the unit on for a few seconds. If the unit detects a problem, the Warning Indicator will be
illuminated. Otherwise, turn the unit off.
• When necessary, clean the Stabilization Strap by removing both strap clips by loosening
the Velcro that holds the clips in place. Machine wash the Stabilization Strap and air dry (do
not tumble dry). Replace both strap clips after the strap has been cleaned and return the
Stabilization Strap to the ACC’s Carrying Case. Otherwise, discard the Stabilization Strap
and replace it with a new one.
• When necessary, clean the Patient Wrist Straps by detaching them from the RMU-1000
ACC and loosening all Velcro. Machine wash and air dry (do not tumble dry). When dry,
Defibtech recommends re-affixing the Patient Wrist Straps to the ACC’s Frame (see
Section 3.5). If the Patient Wrist Straps cannot be cleaned, are damaged, or the Velcro is
worn out, dispose of them and replace with new ones.
Please note that none of the items provided with the RMU-1000 ACC are sterile or require sterilization.
WARNING
Do not sterilize the RMU-1000 ACC or its accessories.
5.3 Storage
The RMU-1000 ACC should be stored in its Carrying Case and placed in a readily accessible
location. In general, the unit should be stored in clean, dry and moderate temperature conditions.
Make sure that the environmental conditions of the storage location are within the ranges detailed
in the “Environmental” section in Chapter 7 of this manual.
Note:
If stored outside of the Compression Module, a Battery Pack in active mode (see Section
3.8 for details about battery states) should be recharged to full capacity at least every 6 months.
If stored in the Compression Module, the Battery Pack should be recharged to full capacity at
least every 2 months.
5.4 Troubleshooting
The following table lists the symptoms, the possible causes, and the possible solutions for
common problems. Refer to the other sections of the user manual for detailed explanations on
how to address any issues. If the unit continues to be non-functional, refer the unit for servicing.
(Refer to Chapter 10 of this manual for contact information.)
Symptom/Observation
Possible Cause
Corrective Action
ACC will not turn on
Battery Pack is not installed
Install charged Battery Pack (see Section 3.8).
Battery Pack depleted
Charge depleted Battery Pack (see Section 3.9) or
replace depleted Battery Pack with a charged Battery
Pack (see Section 3.8).
Battery Pack malfunction
Replace Battery Pack with a charged Battery Pack (see
Section 3.8).
ACC malfunction
Remove ACC from patient (see Section 4.9) and start
manual chest compressions as soon as possible.
Cleaning (continued)
RAC-E1510EN-DN
44
Содержание RMU-1000
Страница 2: ......
Страница 6: ...RAC E1510EN DN 6...
Страница 8: ...1 2 The Defibtech RMU 1000 ACC 7 6 3 4 5 1 2 13 12 14 15 8 9 10 11 RAC E1510EN DN 8...
Страница 12: ...RAC E1510EN DN 12...
Страница 28: ...RAC E1510EN DN 28...
Страница 50: ...RAC E1510EN DN 50...
Страница 54: ...RAC E1510EN DN 54...
Страница 64: ...RAC E1510EN DN 64...