• The RMU-1000 ACC should not be used adjacent to other equipment and that if adjacent,
the RMU-1000 ACC should be observed to verify normal operation in the configuration in
which it will be used.
• Do not use in the presence of flammable gases or anesthetics. Use care when operating
this device close to oxygen sources (such as bag-valve-mask devices or ventilator tubing).
Turn off gas source or move away from patient, if necessary.
2.3
CAUTIONS:
Conditions, hazards, or unsafe practices that may result in minor personal injury, damage
to the RMU-1000 ACC, or loss of data.
• The Compression Module must be assembled and locked to the Frame for proper operation.
• The Frame must be latched to the Backboard for proper operation.
• A new battery is set to ship mode. Must be activated before use! Activate battery by
charging in unit or charger.
• Follow all Battery Pack labeling instructions. Do not use a Battery Pack after its expiration date.
• Use only Defibtech approved batteries and accessories.
• The ACC must be paused in order to replace a Battery Pack. Failure to do so will require the
user to power up the ACC and reset the start position in order to resume compressions.
• Only use Defibtech accessories to power the ACC from an external power source.
• The Battery Pack must always be installed to operate the ACC from external power.
Without the Battery Pack, the ACC will flash the warning indicator and will not perform
compressions.
• The mains power quality must be that of a typical commercial or hospital environment.
• Defibrillation electrodes and pads should not be in contact with and be clear of the Piston
and other ACC components.
• Make sure other equipment and/or drugs are applicable for use with the ACC. Consult the
equipment's operating instructions.
• Avoid gel on chest. Gel on the chest (e.g. from defibrillation pads or ultrasound) in the
Patient Interface Pad target area may result in movement of the piston. Be sure to remove
any gel before use.
• Defibrillation pads or electrodes must be removed or moved away from the Patient
Interface Pad target area.
• Do not use on open wounds or if visible signs of existing injuries.
• The ACC should be applied to the patient’s bare chest. Remove clothing, undergarments
and jewelry before use.
• If the RMU-1000 ACC becomes hot, prolonged contact with the patient’s skin could result
in skin burns. If necessary, remove the patient’s arms from the Patient Wrist Straps.
• Do not allow the use of Stabilization Strap to delay or prevent other treatment of the
patient. Apply strap as soon as possible after beginning compressions and always before
patient movement.
WARNINGS (continued)
2.
W
arnings and Cautions
RAC-E1510EN-DN
15
Содержание RMU-1000
Страница 2: ......
Страница 6: ...RAC E1510EN DN 6...
Страница 8: ...1 2 The Defibtech RMU 1000 ACC 7 6 3 4 5 1 2 13 12 14 15 8 9 10 11 RAC E1510EN DN 8...
Страница 12: ...RAC E1510EN DN 12...
Страница 28: ...RAC E1510EN DN 28...
Страница 50: ...RAC E1510EN DN 50...
Страница 54: ...RAC E1510EN DN 54...
Страница 64: ...RAC E1510EN DN 64...