venTest OPERATORS MANUAL
1.2
Volume Measuring with Pressure Triggering
Volume measurement with 2 pressure
trigger thresholds, P1 and P2.
Volume measurement is going to start with
pressure exceeding P1, ending with
pressure being below P2.
In volume calculation an existing flow-base
will be added.
1.3
Volume Measuring with Pressure/Flow Triggering
Volume measurement with a flow trigger
threshold F1 and a pressure trigger
threshold P2.
Volume measurement is going to start with
flow exceeding F1, ending with flow being
below P2.
In volume calculation an existing flow-base
will be added.
Appendix A
#
Page 79
Содержание venTest
Страница 1: ...Operating Manual venTest Ventilator Tester MN 108 ...
Страница 2: ......
Страница 4: ......
Страница 8: ...Page iv ...
Страница 32: ...venTest OPERATORS MANUAL Overview Chapter 2 Page 18 ...
Страница 116: ...venTest OPERATORS MANUAL Appendix C Page 102 ...
Страница 117: ......