venTest OPERATORS MANUAL
1.4
Trigger Delay with Flow/Pressure Triggering
If the flow/pressure after start falls again
below trigger threshold 1 (F1/P1) within T1
time frame, the volume measurement will be
aborted.
If the flow/pressure after start falls below
trigger threshold 1 (F1/P1) within T1 time
frame, the volume measurement will be
terminated. Otherwise the volume
measurement will be terminated if trigger
threshold 2 (F2/P2) becomes true.
Flow/pressure must be above trigger
threshold (F1/P1) for time T1 minimum to
accept volume measurement.
If within time T2 flow/pressure exceeds
trigger threshold 2 (F2/P2) volume
measurement will not be terminated.
Volume measurement will be terminated if
flow/pressure is below trigger threshold 2
(F2/P2) for time T2 minimum.
Appendix A
#
Page 80