venTest OPERATORS MANUAL
2
4.14.2 O -Sensor (optional)
Once the sensor is connected correctly to the base unit, it is automatically recognized by the
device and is ready for use. In the status bar, the sensor is not displayed.
4.14.3 IRMA™ AX+ (optional)
Connect the IRMA™ sensor to the rear DB-9/RS-232 connector on the rear of the base unit.
To set IRMA™ in operation, the supply voltage must be switched on. Once the power has been
turned on, IRMA™ connects to the device and indicates this by flashing LED's. Once a green LED
lights up steadily, the IRMA™ is connected.
4.15 Troubleshooting & Error Handling
4.15.1 BiFlow-Sensor Error Messages
DISPLAY MESSAGE
ERROR DESCRIPTION /
TROUBLE SHOOTING
Flow sensor defective
BiFlow-Sensor was detected as
defective.
Disconnect sensor and re-connect
again.
Check sensor for contamination.
If the error still occurs, send sensor
in for repair
Please return BiFlow for calibration
BiFlow-Sensor has exceeded its
recommended calibration interval.
Send BiFlow-Sensor in for
calibration
Operation/Chapter 4
#
Page 71