venTest OPERATORS MANUAL
Range and Accuracy
Parameter
Units
Range
Accuracy
(whichever is
greater)
Insp. Peak Flow
l/min; ml/min
0 - 300 L/min (AIR)
2
0 - 200 L/min (O ,
2
AIR/O
± 2% of m.v. or 0.1
l/min (AIR)
‡
± 4% of m.v. or 0.2
2
2
l/min (O , AIR/O )
‡
Exp. Peak Flow
l/min; ml/min
0 - 300 L/min (AIR)
2
0 - 200 L/min (O ,
2
AIR/O
± 2% of m.v. or 0.1
l/min (AIR)
‡
± 4% of m.v. or 0.2
2
2
l/min (O , AIR/O )
‡
2
HFO Pmean
kpa; mbar; cmH O;
mmHg; psi; hpa
0 - 10 kPa
± 2% of m.v. or
0.04 kPa
† ‡
2
HFO PMax
kpa; mbar; cmH O;
mmHg; psi; hpa
0 - 10 kPa
± 2% of m.v. or
0.04 kPa
† ‡
2
HFO PMin
kpa; mbar; cmH O;
mmHg; psi; hpa
0 - 10 kPa
± 2% of m.v. or
0.04 kPa
† ‡
2
HFO amplitude
kpa; mbar; cmH O;
mmHg; psi; hpa
0 - 10 kPa
± 2% of m.v. or 0.04
kPa
† ‡
HFO frequency
Hz
0 - 20
± 1 Hz
HFO I/E
3/1 - 1/3
± 0.3
(†)
Please note that due to the pneumatic resistance of the Flowsense 012 Sensor the measured pressure from the ventilation pressure sensor may
differ from values measured at other places in breathing system, especially at high flows .
(‡)
Subject to trigger settings
Specifications/Chapter 1
#
Page 6