venTest OPERATORS MANUAL
4.15.2 IRMA™ AX+ Sensor Error Messages
DISPLAY MESSAGE
ERROR DESCRIPTION /
TROUBLE SHOOTING
IRMA™ Adapter error
Check correct installation of
adapter.
Adapter may be replaced if error
still occurs.
2
O value of IRMA™ is out of
measurement range!
Please check concentration
2
N O value of IRMA™ is out of
measurement range!
Please check concentration
PA value of IRMA™ is out of
measurement range!
Please check concentration
Pressure value of IRMA™ is out of
measurement range!
Please check applied pressure
Following error occurred on
IRMA™:
Error 1!
A software error occurred. Please
restart sensor.
Following error occurred on
IRMA™:
Error 2!
A hardware error occurred.
Please restart sensor.
If the error still occurs, the sensor
has to be replaced.
Following error occurred on
IRMA™:
Error 4!
A motor error occurred.
Please restart sensor.
If the error still occurs, the sensor
has to be replaced.
Following error occurred on
IRMA™:
Error 8!
The motor speed exceeds limits
Please restart sensor.
If the error still occurs, the sensor
has to be replaced.
Operation/Chapter 4
#
Page 72