venTest OPERATORS MANUAL
4.7 Filter
With this setting only a software filter is applied.
without Filter
Measured values are displayed
unfiltered.
with Filter
Measured values are displayed
filtered, therefore measured values are
more constant.
For the numeric values, an IIR-filter with a cutoff frequency of 1 Hz is used. Unfiltered this data is
brought to the display approximately every 320 ms for the numeric values.
4.8 FlowSense Settings Button
Pressing the FlowSense icon leads directly to the menu of connected FlowSense Sensors in sub
menu 'Sensors'. On this screen, a number of additional settings, such as invert flow, # samples to
average, Insp. and Exp. Trigger levels can be selected. In addition, the ability to reset the flow
and pressure zero point is available under the ‘Adjustment’ selection.
IMPORTANT:
When first received, some zero flow variation may be observed. It
is recommended that a zero point adjustment be performed before
first use. See section 4.12.1.
Operation/Chapter 4
#
Page 41