venTest OPERATORS MANUAL
2
4.12.3 O -Sensor (optional)
This section describes the adjustment of the optional oxygen sensor.
2
The calibration of the O -Sensor is based on the concentration points 21% and 100% oxygen.
After selection of the desired concentration point a confirmation message is displayed to confirm
the accepted value.
Adjustments must be made after providing the correct oxygen concentration (21% ambient air or
100% oxygen) to the sensor.
The actual value measured by the sensor appears at the top right and can be used as an indicator
of whether the measuring value has stabilized at each selected concentration. It should be noted
here that the absolute oxygen value in the calibration is not important, but whether the measuring
value is stabilized.
During calibration the sensor should not be exposed to pressure and the sensor must be held
firmly, in a vertical position, with the sensor opening directed downwards.
Failure to observe these instructions can result in misalignment of the sensor.
Figure 35
Operation/Chapter 4
#
Page 66