venTest OPERATORS MANUAL
4.12.2.3 'Adjustment' Button
This menu item is used for setting the zero level of the connected flow sensor.
Flow and pressure zero values can be set independently.
If the zero level needs to be set, ensure that there is no flow or pressure applied to the sensor.
The FlowSense Sensor shall be located on a solid surface and one side shall be closed by the
cap delivered with the sensor.
‘Adjustment’
expects the sensor being exposed to ambient air. In case of any doubts, the sensor
should be flushed with ambient air for a few seconds.
After confirming the above conditions, the adjustment process can be started by pressing the
desired button (
‘Flow’
or
‘Pressure’
).
Simultaneous illumination of both direction indicators of the FlowSense sensor indicates an active
adjustment is achieved.
If no FlowSense sensor is connected, an error message appears. After re-connection of the
sensor, the process must be restarted.
Figure 34
Operation/Chapter 4
#
Page 64