venTest OPERATORS MANUAL
4.16 Cleaning
Before cleaning, turn off the device and disconnect the electrical power / mains plug.
The following care instructions will be explained in the subsequent paragraphs.
•
Case
•
Display
•
FlowSense Sensor
4.16.1 Housing
The case can be cleaned with a soft cloth and a mild soap solution.
Never allow fluids to penetrate the device or the connections.
4.16.2 Display
The display should be cleaned with a soft cloth and pure water only. The use of microfiber cloths
and chemical cleaning agents is not recommended.
4.16.3 FlowSense Sensor
The flow sensors have a measuring wire per direction that can be damaged by improper cleaning.
Therefore, the following information is essential:
•
The case can be cleaned with a soft cloth and a mild soap solution.
•
Never allow fluids to penetrate the device or the connections.
•
Never clean the sensor with high pressure or under running water.
•
Never clean the sensor in an automated disinfection device or ultrasonic bath.
•
Never autoclave the sensor.
To prevent contamination of the sensor we strongly advise you to cover the ends of the sensor
with the provided protection caps when not in use.
Operation/Chapter 4
#
Page 73