venTest OPERATORS MANUAL
Appendix A: Triggering Options
1
Trigger Options of venTest
Examples of triggering options for the venTest are illustrated in this appendix.
Only volume measurement will be discussed, but frequency and I/E measurement (TI and TE) are
also influenced by trigger conditions.
The flow and pressure diagrams shown are to be regarded as examples as they can take many
other forms in practice (according to the measuring application and instrument under test).
1.1
Volume Measuring with Flow Triggering
Volume measurement with 2 flow trigger
1
thresholds, F and F2.
The volume measurement is going to start
with flow exceeding F1, ending with flow
being below F2.
Volume measurement with a
continuous-flow-base and two flow trigger
thresholds F1 and F2.
Volume measurement is going to start with
flow exceeding F1, ending with flow being
below F2.
Appendix A
#
Page 78