venTest OPERATORS MANUAL
2.1.4 Oxygen Sensor Connection (option)
To connect the optional Oxygen Sensor, line up the pins to the cable connector and then hand
tighten the knurled nut. When disconnecting, ensure the knurled nut is fully disengaged before
pulling gently on the connector.
Note:
The Oxygen sensor option must be activated before it can be identified by venTest. For
pricing and activation information please contact Datrend (see section 4.17.4)
WARNING:
Do not force the connector on or off, or pull on the sensor cable.
Permanent damage to the cable and connector may occur.
2.1.5 IRMA Sensor Connection (option)
TM
To connect the optional IRMA AX+ Multi-gas Sensor for measuring anesthesia gases, connect the
sensor to the DB9/RS232 connector using the cable adapter (7200-241), as shown in Figure 3.
Note:
The IRMA sensor option must be activated before it can be identified by venTest. For
pricing and activation information please contact Datrend (see section 4.17.4)
WARNING:
Do not force the connector on or off, or pull on the sensor cable.
Permanent damage to the cable and connector may occur.
The
IRMA
sensor must be turned on in the Sensor->IRMA screen before use, and is ready for
TM
use if the green LED on the sensor is illuminated. A corresponding symbol will be shown on the
status bar of the LCD display when the sensor is connected correctly. For correct readings, an
airway adapter (7050-070 adults/ped, or 7050-071 infant) must be used. It may be necessary to
do a zero adjustment before use.
Figure 3
Overview/Chapter 2
#
Page 16