venTest OPERATORS MANUAL
2
Average O -
Concentration
Time until
accuracy is
decreased by
approx. 0.5%
(adjustment
interval)
100%
approx. 1 Week
40%
approx. 2 Weeks
20%
approx. 4 Weeks
ATTENTION:
2
The average O concentration is the median oxygen concentration which is
given to the sensor in the course of a day or a week .
2
2
Example: You measure 100% O for 4 hours per day, 21% O for 20 hours. That implies an
average concentration of (4 * 100 % + 20 * 21%)/24 = 34 %
The stated accuracy of the oxygen measurements only apply if calibrations have been carried out
properly.
Please note that the sensor reacts on overpressure (related to the ambient pressure during
calibration adjustment) with a relative error of +0.1% per mbar (Attention: not Vol %!!). At a
measuring value of 50 Vol % and an overpressure of 10 mbar, an error of +0.5 Vol % is
generated, this means that 50.5 Vol% is displayed.
This effect is also present as far as calibration is concerned. Therefore an additional calibration
beyond calibration intervals is necessary if the ambient pressure differs considerably from the
pressure which had existed at the time of calibration.
Operation/Chapter 4
#
Page 75