venTest OPERATORS MANUAL
4.12.1 Zeroing of Pressure Sensors
This sub menu is used to set zero points for all internal pressure sensors of the base unit.
Setting zero points of single sensors is supported as well as setting the zero point of all existing
sensors together.
During this process no pressure must be applied to any of the sensors. All pressure connections
of the venTest shall be left open (no hoses shall be connected).
After confirming that no pressure is applied to the sensors, press the
'Start'
button to enter the
zero adjustment process. Start has to be confirmed by pressing
'Yes'
or
'No'
.
If
'Yes'
is selected, a message appears after successful zero adjustment. This message must be
confirmed.
In the lower part of the screen the last factory calibration date of the sensors is displayed.
Figure 30
Operation/Chapter 4
#
Page 58