Dakota Ultrasonics
42
3) Press the
ENTER
key to display the
Digits Edit Box
.
4) Press the
UP
and
DOWN
arrow keys to scroll the highlighted value.
5) Press the
LEFT
and
RIGHT
arrow keys to scroll the digit locations.
6) Repeat steps
4 & 5
until the known thickness value is correctly displayed.
7) Press the
OK
key to calculate the velocity and return to the menu screen, or
ESC
to cancel the one point calibration.
Apply a drop of couplant on the transducer and place the transducer in steady
contact with the 4” calibration standard. Be sure that the reading is stable and
the repeatability indicator, in the top left corner of the display, is fully lit and