DFX-7
Ultrasonic Flaw Detector
13
4) Press the
ENTER
key to display the
Digits Edit Box
.
5) Press the
UP
and
DOWN
arrow keys to scroll the highlighted value.
6) Press the
LEFT
and
RIGHT
arrow keys to scroll the digit locations.
7) Repeat steps
5 & 6
until the known thickness value is correctly displayed.
8) Press the
OK
key to calculate the velocity and return to the menu screen, or
ESC
to cancel the one point calibration.
9) Finally, press the
MEAS
key to return to the measurement screen and begin
taking readings.
Note: CHECK YOUR CALIBRATION!
Place the transducer back on the
calibration point. The thickness reading should now match the known
thickness. If the thickness is not correct, repeat the steps above.
2.5 Zero Coating
In order to account for very slight electronic differences in transducers of the same
type, frequency, and diameter, the
DFX-7
has been equipped with a “zero coating”
feature. This enables the
DFX-7
to obtain very accurate readings on coatings,
eliminating potential errors incurred from slight differences in the manufacturing
processes. The procedure is outlined below: