46
3) Press the
UP
and
DOWN
arrow keys to toggle the coating option on/off.
4) Wipe all couplant from the transducer face and proceed to the Probe Zero
section that follows.
In this second example the transducer was not identified and will force the user to
select the transducer type from a predefined list of transducers:
Selecting the Transducer Type
5) Press the
OK
or
ESC
keys to display the factory list of transducer types (by
diameter and frequency).