DFX-7
Ultrasonic Flaw Detector
19
8) Press the
OK
key to calculate the coating velocity and return to the menu
screen, or
ESC
to cancel the one point calibration.
9) Finally, press the
MEAS
key to return to the measurement screen and begin
taking readings.
Note: CHECK YOUR CALIBRATION!
Place the transducer back on the
calibration point. The coating thickness reading should now match the known
coating thickness sample. If the thickness is not correct, repeat the steps
above.
2.7 Measure
The
DFX-7
is now ready to measure. There are four different measurement view
options, each with a specific purpose – Digits, RF, RECT, & B-Scan. The steps
below outline how to toggle between the different view mode options:
Selecting the Measurement View Option
1) Press the
MENU
key once to activate the menu items tab. Press the
MENU
key multiple times to tab right and the
ESC
key multiple times to tab left until
the
DISP
menu is highlighted and displaying the submenu items.
Содержание DFX-7
Страница 35: ...Dakota Ultrasonics 34 Note Sets where the detection will occur on the negative or positive going echoes...
Страница 219: ...Dakota Ultrasonics 82 10 Finally press the MEAS key to return to the measurement screen and begin taking readings...
Страница 308: ...171 APPENDIX B SETUP LIBRARY Num Name Comment 1 Gn AGC Velocity 1 Enter Custom Name 2 3 4 5 6...