71
CHAPTER TEN
ADDITIONAL FEATURES OF THE DFX-7
10.1 Material Velocity
The
DFX-7
has a built in feature to manually adjust the material velocity to a know
value in the calibration section of the
DFX-7
. It should be noted that this does not
calculate the zero value like the two point calibration option does. Therefore, if the
material velocity is changed you should check the calibration. The procedures for
adjusting the velocity are outlined below:
Adjusting Velocity
1) Press the
MENU
key once to activate the menu items tab. Press the
MENU
key multiple times to tab right and the
ESC
key multiple times to tab left until
the
CAL
menu is highlighted and displaying the submenu items.
2) Use the
UP
and
DOWN
arrow keys to scroll through the sub menu items
until
VELCITY
is highlighted.
3) Press the
LEFT
and
RIGHT
arrow keys to scroll the value. When the
correct velocity value is being displayed,
proceed to step 8
.
4)
Alternatively
, press the
ENTER
key to display the
Digits Edit Box
.
5) Press the
UP
and
DOWN
arrow keys to scroll the highlighted value.