Performing a Coating Zero
1) Press the
MULTI MODE
key once to activate the measurement mode
options.
2) Use the
UP
and
DOWN
arrow keys to scroll through the sub menu items
until
Coating Only (CT)
mode is highlighted.
3) Press the
ENTER
key to select the measurement mode and return to the
measurement screen.
4) Apply a drop of couplant on the transducer and place the transducer in
steady contact with the probe zero disk (battery cover) and obtain a steady
reading.
Note: The coating measurement displayed will potentially be a value greater or
less than 0.
5) Press the
MENU
key once to activate the menu items tab. Press the
MENU
key multiple times to tab right and the
ESC
key multiple times to tab left until
the
PRB
menu is highlighted and displaying the submenu items.