Dakota Ultrasonics
92
4) Once the desired
FILTERS
is displayed, press the
MEAS
key to return to
the measurement screen.
10.18 Peak Hold
The
DFX-7
includes a peak hold feature that draws and saves the maximum
waveform amplitude on the screen. This allows an inspector to move the transducer
in all directions, in an effort to “peak up” on a signal, and locate the maximum
amplitude. The following procure outlines the steps to activate and use the
PEAK
hold feature:
Activating Peak
1) Press the
MULTI MODE
key located in the bottom left corner of the keypad
to display the mode options.
2) Use the
UP
and
DOWN
arrow keys
PEAK
is highlighted.
3) Press the
MEAS
key to return to the measurement screen.
Clearing Current Wave Draw
Note:
At any point during a
PEAK
scan the operator may opt to clear the wave
drawing and start over.