Dakota Ultrasonics
20
purpose of the design, was to provide the user with an efficient way to make
adjustments on the fly during inspections. The following procedure outlines the steps
to navigate and make adjustments as follows:
Hot Menu
Gate1
1) Press the
MEAS
key once to activate the
Hot Menu
. Press the
MEAS
key
multiple times to advance the cell right, and the
ESC
key multiple times to
advance the cell left, until the
WIDTH1
cell is highlighted.
2) Press the
LEFT
and
RIGHT
arrow keys to toggle the value or status.
Note:
This is a coarse adjustment. It’s the fastest and easiest way to
increment, decrement, or toggle the status, however it does so in coarse steps.
Set Detect
1) Press the
UP
and
DOWN
arrow keys until
DETECT
is highlighted.
2) Press the
LEFT
and
RIGHT
arrow keys to toggle the
ABOVE
or
BELOW
.
Note:
Above will trigger the alarm when a signal meets or exceeds the height
of the gate, and below will detect on anything that meets or is below the height
of the gate.
3)
Alternatively
, press the
ENTER
key to display the
Digits Edit Box
.
4) Press the
UP
and
DOWN
arrow keys to scroll the highlighted value.
5) Press the
LEFT
and
RIGHT
arrow keys to scroll the digit locations.