DFX-7
Ultrasonic Flaw Detector
127
Note:
Before toggling the
Polarity
, the
DFX-7
should be set to the
RF
display
view option. The
RF
view will give the user the best opportunity to correctly
view the positive and negative cycles of the waveform. Please refer to page 61
for information on selecting the
Display Views
.
1) Press the
MENU
key once to activate the menu items tab. Press the
MENU
key multiple times to tab right, and the
ESC
key multiple times to tab left,
until the
TUNE
menu is highlighted and displaying the submenu items.
2) Use the
UP
and
DOWN
arrow keys to scroll through the sub menu items
until
POLARITY
is highlighted.
3) Use the
LEFT
and
RIGHT
arrow keys to toggle the
POLARITY
negative or
positive.
4) Press the
MEAS
key to return to the measurement screen and begin taking
readings.
Important:
Be sure to do a Probe Zero after changing the polarity!
10.12 Pulse Width
The
DFX-7
has an adjustable pulse width option. Pulse width, refers to the duration
of time the pulser is left on. This time results in increased energy sent into the test