DFX-7
Ultrasonic Flaw Detector
81
1) Press the
MENU
key once to activate the menu items tab. Press the
MENU
key multiple times to tab right, and the
ESC
key multiple times to tab
left, until the
GT1
menu is highlighted and displaying the submenu items.
2) Use the
UP
and
DOWN
arrow keys to scroll through the sub menu items
until
THRESHOLD1
is highlighted.
3) Press the
LEFT
and
RIGHT
arrow keys to scroll the value. When the
correct threshold is being displayed,
proceed to step 8
.
4)
Alternatively
, press the
ENTER
key to display the
Digits Edit Box
.
5) Press the
UP
and
DOWN
arrow keys to scroll the highlighted value.
6) Press the
LEFT
and
RIGHT
arrow keys to scroll the digit locations.
7) Repeat steps
5 & 6
until the
Threshold
number is correctly displayed.
8) Press the
OK
key to set the
Threshold
and return to the menu screen, or
ESC
to cancel entering the
Threshold
.
9) Repeat steps
1-8
to set any additional
Threshold
level.