DFX-7
Ultrasonic Flaw Detector
103
and place the transducer in steady contact with the coating and sample or
actual test material. Be sure that the reading is stable and the repeatability
indicator, in the top left corner of the display, is fully lit and stable. Press the
MENU
key once to activate the menu items tab. Press the
MENU
key
multiple times to tab right and the
ESC
key multiple times to tab left until the
CAL
menu is highlighted and displaying the submenu items.
3) Use the
UP
and
DOWN
arrow keys to scroll through the sub menu items
until
COATING 1PT
is highlighted.
4) Press the
ENTER
key to display the
Digits Edit Box
.
5) Press the
UP
and
DOWN
arrow keys to scroll the highlighted value.
6) Press the
LEFT
and
RIGHT
arrow keys to scroll the digit locations.
7) Repeat steps
5 & 6
until the known thickness value is correctly displayed.
8) Press the
OK
key to calculate the velocity and return to the menu screen, or
ESC
to cancel the coating one point calibration.
9) Finally, press the
MEAS
key to return to the measurement screen and begin
taking readings.