DATA ANALYSIS
Table 5 CBC/DIFF Parameter Codes
On DMS Display
Cause
continued
"-----" for CBC parameter results, and no average
histogram for the affected parameter
If WBC is "-----", then LY#, MO#, NE#, EO#,
and BA# are "....."
Total voteout (at least 2 out of 3 measurements do
not agree). Diff parameters do not vote out.
"+++++" for parameter results
IMPORTANT
If the WBC, RBC, Hgb or Plt result is +++++, to
prevent carryover to the next sample, run a
blank cycle.
Result exceeds printable range:
RBC > 10.00 x 10
6
Plt > 1000. x 10
3
MCV > 1000.
WBC > 100.0 x 10
3
If WBC is overrange
an R appears next to the RBC, Hgb, Hct,
MCV, MCH, MCHC, RDW, MPV, Plt and
Diff% results. Results for Diff # are dots (....).
If RBC is overrange
an R appears next to Hct, MCH and MCHC
results.
If Hgb is overrange
an R appears next to MCH and MCHC
results.
If MCV is overrange
an R appears next to Hct, MCHC and RDW
results.
If Plt is overrange
an R appears next to WBC, Hgb, MCH,
MCHC, diff number and Pct results.
Average values for overrange WBC, RBC or Plt
are displayed in Data Base Query,
Edit,
beneath the Definitive flag section. Press
to
see them. These values are provided for review
only; do not report them.
continued
80
PN 4237188A (December 1993)
Содержание STKS
Страница 16: ...CONTROLS AND INDICATORS 6 PN 4237188A December 1993...
Страница 84: ...SAMPLE ANALYSIS 74 PN 4237188A December 1993...
Страница 98: ...DATA ANALYSIS 88 PN 4237188A December 1993...
Страница 124: ...DMS BASICS 114 PN 4237188A December 1993...
Страница 130: ...RUN SAMPLES DISPLAY 120 PN 4237188A December 1993...
Страница 168: ...DATA BASE 158 PN 4237188A December 1993...
Страница 204: ...APPENDIX A 194 PN 4237188A December 1993...
Страница 216: ...APPENDIX C 206 PN 4237188A December 1993...
Страница 228: ...APPENDIX D 218 PN 4237188B March 1995...
Страница 230: ...REFERENCES 220 PN 4237188B March 1995...
Страница 236: ......