
STARTUP
6.
Mix controls gently according to directions
in the package insert. Perform the 8/8/8
procedure twice.
8 TURNS
7022034B
8 TURNS
8 INVERSIONS
7.
Cycle the control:
a.
Immerse the aspirator tip in the control
vial.
b.
Press and release the activator.
c.
Remove the control when
•
you hear the beep and
•
the Analyzer CRT status field
displays
WIPE
.
d.
Wipe the tip with a lint-free tissue.
7188057A
8.
View the results on the CBC/DIFF
CONTROL RUN screen. To see WBC, RBC
and Hgb graphs of the last 10 cycles, press
.
DMS
Press
again to view the other parameter
graphs.
9.
To review control files, control graphics and
graphs, see the Controls chapter.
40
PN 4237188A (December 1993)
Содержание STKS
Страница 16: ...CONTROLS AND INDICATORS 6 PN 4237188A December 1993...
Страница 84: ...SAMPLE ANALYSIS 74 PN 4237188A December 1993...
Страница 98: ...DATA ANALYSIS 88 PN 4237188A December 1993...
Страница 124: ...DMS BASICS 114 PN 4237188A December 1993...
Страница 130: ...RUN SAMPLES DISPLAY 120 PN 4237188A December 1993...
Страница 168: ...DATA BASE 158 PN 4237188A December 1993...
Страница 204: ...APPENDIX A 194 PN 4237188A December 1993...
Страница 216: ...APPENDIX C 206 PN 4237188A December 1993...
Страница 228: ...APPENDIX D 218 PN 4237188B March 1995...
Страница 230: ...REFERENCES 220 PN 4237188B March 1995...
Страница 236: ......