STARTUP
19. Wait 30 seconds then analyze the prepared
sample:
a.
Immerse the aspirator tip in the
dilution.
b.
Press and release the activator.
IMPORTANT
Do not remove the sample tube from the
aspirator tip until the STATUS message on the
Analyzer CRT changes from INTRO SAMPLE to
WIPE, and there is a beep.
c.
Remove the prepared control when
•
you hear the beep and
•
the Analyzer CRT STATUS field
displays WIPE.
d.
Wipe the tip with a lint-free tissue.
7188037A
20. Check the results on the RETIC CONTROL
RUN screen.
RETIC CONTROL RUN
FILE:
18
LOT: 539900
OPR: hs
SHIFT : 0
LEVEL: Level III
EXP DATE: 11/26/93
RUN:
22
DATE: 10/12/93
TIME: 10:08
Ref RBC
4.65
RESULTS
ASSAY
DIFF
LIMIT
RET%
9.22
9.40
-.180
1.50
COMPUTED
COMPUTED
RESULTS
VALUE
DIFF
LIMIT
RET#
.4289
.4371
-.0082
.0698
F1-Help
F2-File
F4-Print
F6-Graph
F8-Delete
F9-Menu
F12-Graphics
11/15/93 14 59
OPR
DMS
↑
PR
↓
TK
↓
HC
↓
DB
↑
XB
↑
WL
↑
HWL
↓
QC
↓
53
3
To see the RET% and RET# graphs of the
last 10 cycles, press
Graph
.
DMS
32
PN 4237188A (December 1993)
Содержание STKS
Страница 16: ...CONTROLS AND INDICATORS 6 PN 4237188A December 1993...
Страница 84: ...SAMPLE ANALYSIS 74 PN 4237188A December 1993...
Страница 98: ...DATA ANALYSIS 88 PN 4237188A December 1993...
Страница 124: ...DMS BASICS 114 PN 4237188A December 1993...
Страница 130: ...RUN SAMPLES DISPLAY 120 PN 4237188A December 1993...
Страница 168: ...DATA BASE 158 PN 4237188A December 1993...
Страница 204: ...APPENDIX A 194 PN 4237188A December 1993...
Страница 216: ...APPENDIX C 206 PN 4237188A December 1993...
Страница 228: ...APPENDIX D 218 PN 4237188B March 1995...
Страница 230: ...REFERENCES 220 PN 4237188B March 1995...
Страница 236: ......