STARTUP
CHECK STARTUP RESULTS
1.
Look at the START UP screen when results
appear on the DMS. Check background
results. The preset tolerances are:
WBC
0.20
RBC
0.01
Hgb
0.15
Plt
3.00
Out-of-limit background results are in red
and are flagged H (High) or L (Low).
If an incomplete computation (...) of the
Hgb parameter occurs, verify that the Hgb-
blank and Hgb-read voltages are acceptable
according to the guidelines in step 2. If the
step 2 voltages are acceptable, the dots (...)
for the Hgb parameter are acceptable.
If the background results are not acceptable,
go to step 3.
START UP
REAGENT DATES 08/20/93
RAMP TEST 08/20/93
PRECISION TEST
08/20/93
Diluent 12/22/93
WBC
50.56
WBC
63.05
Lyse
11/22/93
RBC
5.062
RBC
6.384
Cleaner 12/22/93
HGB
14.20
HGB
14.20
Pak
11/22/93
HCT
83.83
HCT
47.17
MCV
165.5
MCV
73.8
DIFF TEST 08/20/93
MCH
28.06
MCH
22.25
MCHC 16.94
MCHC 30.12
Diff Pressure
0.37
RDW
47.67
RDW
14.23
Sheath Pressure
6.38
PLT
276.4
PLT
798.7
Laser Voltage
5.09
PCT
.3151
PCT
.4784
RF Voltage
6.34
MPV
11.39
MPV
5.98
PDW
18.24
PDW
16.45
HGB VOLTAGE CHECK 08/20/93
BACKGROUND TEST 08/20/93
HGB Blank
8.89
WBC
.12
HGB
.02
HGB Read
8.88
RBC
.001
PLT
.0
F1-Help
F4-Print
F9-Menu
08/20/93 10:29
OPR
DMS
↑
PR
↑
TK
↑
HC
↓
DB
↑
XB
↓
WL
↑
HWL
↓
QC
↑
132
1
2.
Verify that the Hgb readings are within
limits. The criteria for the Hgb readings are:
•
The Hgb-blank and Hgb-read voltages
are within 7.500 to 9.500.
•
The Hgb-blank voltage is within
±0.075 V of the Hgb-read voltage.
•
The Hgb-blank voltage is within
±0.150 V of the Hgb-blank voltage for
the previous day.
•
The Hgb-read voltage is within
±0.150 V of the Hgb-read voltage for
the previous day.
If the readings do not meet these criteria, go
to step 3.
16
PN 4237188A (December 1993)
Содержание STKS
Страница 16: ...CONTROLS AND INDICATORS 6 PN 4237188A December 1993...
Страница 84: ...SAMPLE ANALYSIS 74 PN 4237188A December 1993...
Страница 98: ...DATA ANALYSIS 88 PN 4237188A December 1993...
Страница 124: ...DMS BASICS 114 PN 4237188A December 1993...
Страница 130: ...RUN SAMPLES DISPLAY 120 PN 4237188A December 1993...
Страница 168: ...DATA BASE 158 PN 4237188A December 1993...
Страница 204: ...APPENDIX A 194 PN 4237188A December 1993...
Страница 216: ...APPENDIX C 206 PN 4237188A December 1993...
Страница 228: ...APPENDIX D 218 PN 4237188B March 1995...
Страница 230: ...REFERENCES 220 PN 4237188B March 1995...
Страница 236: ......