STARTUP
11. For each level of control, label two glass test
tubes: one for Reagent A and one for
Reagent B. Mark the control level on both
tubes.
B
Ι
A
Ι
7170001A
IMPORTANT
Dispensing reagent at an angle changes the
dilution of the preparation.
12. Holding the dropper bottle vertically, place 4
drops of Reagent A into the "A" tube.
7170002A
A
Ι
A
13. Dispense 50 µL of well-mixed Retic-C cell
control into the "A" tube. Add directly to the
stain and do not let control run down side of
tube.
7170003A
50 µL
A
Ι
14. Gently mix "A" tube.
A
7170050A
30
PN 4237188A (December 1993)